MODEL | 7730 |
AC WITHSTAND VOLTAGE |
| Range | Resolution | Accuracy |
Output Voltage ACV | 0-5000 | 1 | ±(2% of setting+ 5V) |
Output Frequency | 50Hz/60Hz ±0.1% User Selection |
Output Waveform | Sine Wave THD: Less than 2% at 300Vrms or higher voltage (Resistive Load) CrestFactor = 1.3 - 1.5 |
Output Regulation | ± (1% of output + 5V) From no load to full load |
SETTINGS |
Maximum Limit current mA | 0.000 - 9.999 (0=OFF) | 0.001 | ± (2% of setting + 2 counts) |
10.00-30.00 | 0.01 |
Minimum Limit current mA | 0.000 - 9.999 (0=OFF) | 0.001 | ± (2% of setting + 2 counts) |
10.00-30.00 | 0.01 |
Ramp-Up Timer second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) |
Ramp-Down Timer second | 0.0 - 999.9 |
Dwell Timer second | 0 0.4 - 999.9 |
(0=continuous) |
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) |
DC WITHSTAND VOLTAGE |
Output Voltage DCV | 0-5000 | 1 | ±(2% of setting+ 5V) |
Output Ripple | <4 %(5KV/10mA at Resistive Load) |
SETTINGS |
Maximum Limit current µA | 0 - 999.9 | 0.1 | ± (2% of setting + 2 counts) |
1000 - 10000 | 1 |
Minimum Limit current µA | 0.0 - 999.9 | 0.1 | ± (2% of setting + 2 counts) |
1000 - 10000 | 1 |
Ramp-Up Timer second | 0.1 - 999.9 | 0.1 | ± (0.1% + 0.05 sec) |
Ramp-Down Timer second | 0 1.0 - 999.9 (0=OFF) |
Dwell Timer second | 0 0.4 - 999.9 |
(0=continuous) |
Ramp-HI current | >12 mA peak maximum ON/OFF User Selection |
Charge-LO current | 0.0 - 350.0 µA DC or Auto Set |
Discharge Time | ≦ 200 msec |
Maximum Capacitive LoadDC Mode | 1 µF < 1KV 0.08 µF < 4KV |
0.75 µF < 2KV 0.04 µF < 5KV |
0.5 µF < 3KV |
Arc Detection | 1 - 9 ranges (9 is the most sensitivity) |
INSULATION RESISTANCE |
Output Voltage VDC | 50 - 1000 | 1 | ± (2% of setting + 2 counts) |
Short Circuit Current | Maximum >12mA peak |
SETTINGS |
Maximum and Minimun Limit | 0.05 - 99.99(Max – Limit: 0 = OFF) | 0.01 | Same as Resistance MEASUREMENT Accuracy |
Resistance MΩ | 100.0-999.9 | 0.1 |
| 1000-50000 | 1 | ± (0.1% + 0.05 sec) |
Ramp-Up Timer second | 0.1 - 999.9 | 0.1 |
Ramp-Down Timer second | 0 1.0 - 999.9 (0=OFF) |
Delay Timer second | 0 1.0 - 999.9 | 0.1 |
(0=continuous) |
Charge-LO current µA | 0.000 - 3.500 or Auto Set |
MEASUREMENT |
| Range | Resolution | Accuracy |
Voltage KV (AC/DC) | 0.00-5.00 | 0.01 | ± (1.5% of reading +2counts) |
Voltage Vdc (IR only) | 0-1000 | 1 | ± (1.5% of reading + 2counts) |
AC Current mA | 0.000-3.500 | 0.001 | ± (2% of reading + 2counts) |
3.00-30.00 | 0.01 |
DC Current µA | 0.0 - 350.0 | 0.1 | ± (2% of reading + 2 counts) |
DC Current mA | 0.300 - 3.500 | 0.001 |
3.00-9.99 | 0.01 |
Resistance MΩ | 50-499V | 500-1000V | | 50 – 499V |
0.05–999.9 ±(7% of reading+2 counts) |
0.050-1.999 | 0.050-9.999 | 0.001 | 500 – 1000V |
2.00-19.99 | 10.00-99.99 | 0.01 | 0.10–999.9 ±(2% of reading+2 counts) |
20.0-199.9 | 100.0-999.9 | 0.1 | 1000–9999 ±(5% of reading+2 counts) |
200-50000 | 1000-50000 | 1 | 10000–50000 ±(15%of reading+2 counts) |
GENERAL |
Input Voltage | 115/230Vac±15% Auto Selection 50/60Hz ± 5% Fuse 4A Slow-Blo 250VAC |
PLC Remote Control | Input: Test Reset Interlock Recall Memory 1 through 10 |
Output: Pass Fail Test-in-Process Reset Out |
Remote Output Relays | 125VAC@1AAC 30VDC@ 0.5ADC |
Memory | 50 memories 30 steps/memory |
Graphic Display | 240 x 64 dot resolution Monographic LCD /Contrast 9 Levels 1-9 |
Safety | Built-in Smart GFI circuitGFI trip current 450µA max HV shut down speed: <1mS |
Interface | Standard RS232 Optional Printer PortGPIB (IEEE-488.2)Multi-function Interface CardEthernet Card. |
Security | Programmable password lockout capability to avoid unauthorized access to test set-up program. |
Alarm Volume Setting | Range: 0-9 ;0=OFF 1 is softest volume 9 is loudest volume. |
Calibration | Adjustments are made through the fornt panel. Built-in Verification program. |
Automatic Calibration alert function to signal operator when calibration is due. |
Environment | 0-40℃ 20-80%RH |
Dimensions/Net Weight | (W × H × D) 2U (430 × 89 × 400 mm)/13Kg. |
Option (Opt.735) | DC Continuity Test Function (0.1A/ 0-2KΩ) at Rear Panel |
STANDARD ACCESSORIES |
Power Cord(10A) | ×1 |
Fuses | ×2 (Including a spare contained in the fuse holder) |
Interlock Disable Key(1505) | ×1 |
Hipot Test Lead 1.5m(1101) | ×1 |
Hipot Return Lead 1.8m(1102) | ×1 |
RS232 Link Cable 1.5m(1130) | ×1 |
MATRIX SCANNER MODULE (Opt.729) |
High Voltage Rating | 5KVAC/5KVDC |
Number of HV Channel | 8 |
MATRIX SCANNER MODULE (Opt.730) |
High Voltage Rating | 5KVAC/5KVDC |
Number of HV Channel | 4 |
DC CONTINUITY TEST FUNCTION (Opt.735) |
MEASUREMENT |
| Range | Resolution | Accuracy |
Output Current | DC 0.1A ± 0.01A Max. |
Resistance Display | 0.00 – 19.99 | 0.01 | ± (1 % of reading + 5 counts) |
20.0 – 199.9 | 0.1 | ± (1 % of reading + 2 counts) |
200 – 2000 | 1 | ± (1 % of reading + 2 counts) |
SETTINGS |
Max-Limit Ω | 0.00 – 19.99 | 0.01 | ± (1 % of setting + 5 counts) |
20.0 – 199.9 | 0.1 | ± (1 % of setting + 2 counts) |
200 – 2000 | 1 | ± (1 % of setting + 2 counts) |
Min-Limit Ω | 0.00 – 19.99 | 0.01 | ± (1 % of setting + 5 counts) |
20.0 – 199.9 | 0.1 | ± (1 % of setting + 2 counts) |
200 – 2000 | 1 | ± (1 % of setting + 2 counts) |
Dwell Timer second | 0.0 0.3 – 999.9 | 0.1 | ± (0.1% + 0.05 sec) |
(0 = continuous) |
Resistance Offset Ω | 0.00 – 2.00 | 0.01 | ± (1 % of reading + 5 counts) |